Product Details:
Payment & Shipping Terms:
|
Chamber Volume: | 1500 Liter | Door Type: | Automatic Sliding Door |
---|---|---|---|
Chamber: | Rectangle, 304 Stainless Steel | Sterilization Procedure: | Steam |
Cofiguration: | Recessed | Operator Panel: | Touch Screen |
Highlight: | medical waste sterilizer,medical waste treatment equipment |
Recessed Wall Double Door Sterilizers With Water Pump For Water Recycling And Circulation
Quick details
Horizontal sliding door | For stable and easy operation |
Touch screen | Easy to use |
The technical area | On the left or on the right or both sides |
PLC control system | User friendly and one key operation function |
Special design | Customized orders are available |
Applicaitons
The double door autoclaves is designed for central sterile supply departments (CSSD), hospital operating rooms (OR), ambulatory surgery centers, dental schools and other clinical environments.
Specifications
Model |
MD-1.5Y |
Door number |
Double door |
Door type |
Horizontal sliding door |
Chamber volume |
1500 liters |
Chamber Size |
680*1180*1870mm |
Overall Size |
2200*2000*2280mm |
Net Weight |
2230kg |
Steam |
50Kg/cycle |
Water |
150Kg/cycle |
Power(KVA) |
4.5KVA |
Working temperature |
105---134℃ |
Sterilization time |
0---99min |
Temperature difference |
≤±1℃ |
Max working pressure |
0.23MPa |
Steam |
0.4---0.8MPa |
Soft water |
0.1---0.3MPa |
Compressed air |
0.4---0.8MPa |
Electric power |
AC380v, 3P/N/E, 50Hz |
Door
The MD-1.5Y range can be supplied with either single or double (pass-through) doors, and are of the automatic horizontal sliding type. All door configurations include limit switches to detect when the doors are fully open or fully closed, and prevent the doors from being opened until the pressure within the chamber is within safe limits.
Vacuum System
Vacuum is created by means of either a liquid ring vacuum pump, or a water ejector driven by a centrifugal pump. Liquid ring vacuum pumps are fitted with a high efficiency tube condenser between the chamber drain and the pump to prevent overheating and cavitation of the pump, and resulting in a long service life. Each autoclave is fitted with a recirculating water conservation system. This reduces water consumption by the vacuum and condenser system by up to 80%.